Abstract

A gas chromatographic system for the simultaneous acquisition of hydrocarbon and sulfur chromatograms was developed. Detection of sulfur compounds is achieved using a sulfur chemiluminescence detector (SCD) mounted in series with a flame ionization detector (FID). A constant fraction of the effluent of the FID is transferred to the SCD by means of a fixed restrictor. Unlike previous versions of this approach, the FID is not used to generate the chemiluminescent sulfur species. Rather, the FID is operated under optimum conditions for hydrocarbon analysis and a furnace is used to generate the chemiluminescent sulfur species. The system permits dual acquisition of the hydrocarbon and sulfur signals in a single analysis with a single column, since the detectors are operated in a serial fashion. The application of sulfur simulated distillation using this approach was examined, since this requires simultaneous universal and sulfur selective detection. Precision of absolute response of both the FID and SCD was typically less than 2% RSD for a standard reference material.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.