Abstract

In K-edge subtraction (KES) imaging with synchrotron radiation computed tomography (SRCT), two images are taken simultaneously using energies above and below the K-absorption edge of a contrast agent. A logarithmic difference image reveals the contrast agent concentration with good accuracy. Similarly, in temporal subtraction imaging (TSI) the reference image is taken before the introduction of the contrast agent. Quantitative comparisons of in vivo images of rabbit lung indicated that similar results for concentrations of iodine in blood vessels and xenon in airways are obtained by KES and TSI, but the level of noise and artifacts was higher in the latter. A linear fit showed that in the lung parenchyma ρTSI = (0.97 ± 0.03)ρKES + (0.00 ± 0.05) for xenon and ρTSI = (1.21 ± 0.15)ρKES + (0.0 ± 0.1) for iodine. For xenon the calculation of time constant of ventilation gave compatible values for both of the methods. The two methods are combined for the simultaneous determination of the xenon concentration (by KES) and the iodine concentration (by TSI) in lung imaging, which will allow simultaneous in vivo determination of ventilation and perfusion.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.