Abstract

A sensitive ion-pair reversed phase liquid chromatographic method for determining tryptophan, 5-hydroxytryptophan, 5-hydroxytryptamine (serotonin) and 5-hydroxyindole acetic acid has been developed. To separate these indoles in cerebrospinal fluid both octadecyl- and phenyl-bonded columns were tested. The effect of an ion-pair reagent (n-heptyl sulphonate) and pH of the mobile phase on the separation was examined. With the method described the concentrations of the four indoles can be determined within 45 min without sample pretreatment. The sensitivity obtained with electrochemical detection is 0.45 mumol/l for tryptophan and 2-4 nmol/l for 5-hydroxyindoles in human cerebrospinal fluid samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.