Abstract
A method has been developed enabling direct analysis (i.e. after dilution only) of beer, instant coffee, milk, and milk powder by ICP-AES. Analysis of the beverages after dilution with a low concentration of HNO3 was used for accurate determination of essential minor and trace elements (Ca, Mg, K, P, S, and Zn). Selenium, introduced as the hydride, was determined simultaneously with the other non-hydride-forming elements using the commercial multi-mode sample-introduction system (MSIS). To obtain accurate results, however, some simple pre-treatment was needed. Analysis was also performed after microwave-assisted decomposition of the samples. Three different modes of sample-preparation, i.e. dilution only, partial decomposition (aqua regia treatment), and complete decomposition were compared. The results obtained by use of the three different sample-preparation methods were in very good agreement. Results from analysis of certified reference material (SRM 1459 non-fat milk powder) also verified the accuracy of the methods. The limit of detection obtained for Se using dual-mode sample introduction was 0.5 ng mL(-1), which corresponds to approximately 2 ng g(-1) in beer and approximately 4 ng g(-1) in coffee and milk when using the recommended procedure.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.