Abstract

A gas chromatographic procedure for the simultaneous determination of glycerol, mono-, di- and triglycerides in vegetable oil methyl esters has been developed. Quantitative information about this group of organic contaminants is very important for the quality of these oleochemical products when used as automotive diesel fuel substitutes. Trimethylsilylation of glycerol, mono- and diglycerides, followed by GC using a 10-m capillary column coated with a 0.1-μm film of DB-5 allows the determination of all analytes in a single GC run. Calibration is performed by analysis of standard solutions containing glycerol, mono-, di- and triolein as well as two internal standards, 1,2,4-butanetriol and tricaprin. The recovery of the procedure at different concentration levels and the repeatability of the quantitative results are evaluated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.