Abstract

A simple and sensitive high-performance liquid chromatographic (HPLC) method was developed to determine the amount of hydrazine in industrial wastewater after derivatization with anthracene-2,3-dicarbaldehyde. The best reaction conditions (a reagent dosage of 200 mg/L, a pH of 11, and reaction for 1 min at room temperature) were established by the variation of parameters. The formed derivative was injected into the HPLC without further extraction or purification procedures. Under the established conditions, the limits of detection and the limits of quantification were 0.02 µg/L and 0.05 µg/L, respectively, and the inter-day relative standard deviation was less than 9% at concentrations of 0.1 and 0.5 µg/L. The calibration curve showed good linearity with r2 = 0.9999. The method was successfully applied to the determination of hydrazine in various kinds of industrial wastewater.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.