Abstract

The wear of zirconia femoral heads against alumina acetabular inserts with swing-phase microseparation was investigated in a hip joint simulator. Under mild microseparation conditions, the wear was very low, with an average wear rate of 0.05 mm 3/million cycles reported over 5 million cycles of testing. However, under severe microseparation conditions representative of greater joint laxity, the wear rate of zirconia against alumina increased by 2 orders of magnitude, producing severe wear and, in one case, femoral head fracture. The adverse results of this study indicate that the combination of a zirconia femoral head articulating against an alumina acetabular insert is not recommended for clinical use. The results further raise concerns over the suitability of conventional simulators in evaluating the wear of ceramic hip prostheses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.