Primary Serratia marcescens bacteremia developed in 17 patients in an intensive care unit after exposure to pressure monitoring devices. A study showed that all of the transducer heads were contaminated with S marcescens, and prospective culturing of 110 pressure monitoring lines disclosed a 24% rate of contamination with the same organism. Hand contamination occurs at the time the equipment is initially assembled; in five of eight trials, transmission was shown experimentally to occur by direct inoculation of open ports. Routine disinfection of the transducer heads with glutaraldehyde not only effectively decontaminated all pressure monitoring lines in use, but also controlled the outbreak.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.