Abstract
Loop-mediated isothermal amplification (LAMP) holds great potential for point-of-care (POC) diagnostics due to its speed and sensitivity. However, differentiation between spurious amplification and amplification of the target sequence is a challenge. Herein, we develop the use of molecular beacon (MB) probes for the sequence-specific detection of LAMP on commercially available lateral flow immunoassay (LFIA) strips. The detection of three unique DNA sequences, including ORF1a from SARS-CoV-2, is demonstrated. In addition, the method is capable of detecting clinically relevant single-nucleotide polymorphisms (BRAF V600E). For all sequences tested, the LFIA method offers similar sensitivity to fluorescence detection using a qPCR instrument. We also demonstrate the coupling of the method with solid-phase microextraction to enable isolation and detection of the target sequences from human plasma, pond water, and artificial saliva. Lastly, a 3D printed device is designed and implemented to prevent contamination caused by opening the reaction containers after LAMP.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.