Abstract

Stereospecific separations of seven Tic-hydantoin sigma-1 agonists were performed by both HPLC method using derivatized cellulose and amylose chiral stationary phases and capillary electrophoresis (CE) method using neutral and anionic cyclodextrins added in the background electrolyte (BGE). An optimal baseline separation ( R s > 3.3 with analysis times < 25 min) was readily obtained with all silica-based celluloses and amyloses using a normal-phase methodology. CE was used as an alternative technique to HPLC for the Tic-hydantoin derivatives separation. The enantiomers were fully resolved with highly sulfated β-cyclodextrins at pH 2.5 ( R s > 1.5 with analysis times <11 min). Both methods were validated in terms of linearity, detection and quantification limits. They were used to check the enantiomeric purity of the enantiomers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.