This paper investigated the current sensitivity of a resonant sensor based on a magnetically actuated piezoresistive microcantilever in different resonant modes, which were the first flexural mode and the first and second torsional modes. The sensor was based on the idea of measuring the electrothermally induced resonance frequency shift as a result of the Joule heating dissipated when the DC current flowed through the Wheatstone bridge on the microcantilever. Two theoretical models between the microcantilever’s resonance frequency and the square of the applied DC current for the sensor operating under the flexural and torsional modes were established. From the experimental results, it can be seen that the current sensitivity of the first torsional mode is an order of magnitude greater than the first flexural mode, but less than that of the second torsional mode. In addition, the effect of the DC current’s direction on the measured results should be taken into account before detecting the DC current.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.