Abstract
We have presented a simple and sensitive method for determining pethidine, a narcotic analgesic drug in body fluids by gas chromatography–tandem mass spectrometry (GC–MS/MS). Pethidine and 4′-piperidinoacetophenone (internal standard) were extracted from body fluids with Bond Elut C 18 columns; the recoveries were above 85% for both compounds. The calibration curves for blood and urine showed good linearities in the range of 1.25–40 ng/ml. Its detection limits (signal-to-noise ratio=3) were estimated to be approximately 0.5 ng/ml of whole blood and urine.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.