Abstract

A novel and sensitive dispersive liquid-liquid microextraction method based on the solidification of the floating organic drop combined with high-performance liquid chromatography and ultraviolet detection was used for the determination of atorvastatine in blood serum samples. The chromatographic separation of atorvastatin was carried out using methanol as the mobile phase organic modifier. Various parameters affecting the extraction efficiency were optimized, such as the kind and volume of extraction solvent (1-undecanol) and disperser solvent (acetonitrile), pH, and the extraction time. The calibration curve was linear in the range of 0.2-6000 μg/L of atorvastatin (r(2) = 0.995) with a limit of detection of 0.07 μg/L. The relative standard deviation for 100 μg/L of atorvastatin in human plasma was 8.4% (n = 4). The recoveries of plasma samples spiked with atorvastatin were in the range of 98.8-113.8%. The obtained results showed that the proposed method is fast, simple, and reliable for the determination of very low concentrations of atorvastatin in human plasma samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.