Abstract

We correct the values of the group delay dispersion of the cavity mirrors and N2, as well as the concentration of CO2, obtained from the measurement of the center frequencies of cavity modes using a comb-based Fourier transform spectrometer. The corrected values of group delay dispersion are a factor of 3 higher, which implies that the precision and accuracy of the dispersion measurements is 0.3 fs2 and 3 fs2, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.