Abstract

Influence of aging on energy resolution, length of energy resolution plateau, counting characteristic and self-induced space charge effect has been measured. We report results of aging studies performed by irradiating brass sealed proportional wire counters, filled with Ar/CO 2 (90/10), Ar/CO 2 /N 2 /iso-pentane, (70/10/10/10) mixtures. The counters had different anode radii (15, 25 and 50 μ m, respectively). 90 Sr and 109 Cd were used as radiation sources. Our experience shows that the aging rate depends not only on the total collected charge but also on the radius of the anode wire and on the type of radiation source.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.