Abstract

Stress relaxation during pullout of a pedicle screw decreases the peak load and stiffness of the bone-screw interface. However, it is unknown whether this can be generalized to all types of screw designs. This study aimed to show whether screw design altered the effects of stress relaxation on the mechanical performance of the pedicle screw during pullout. Twelve calf vertebrae were obtained: six vertebrae were instrumented with 7.5x40 mm conical pedicle screws and the other six with 5.0x40 mm cylindrical pedicle screws. The screws with two different designs were pulled out using either a standard pullout or a stress relaxation pullout protocol. Both bone-screw interfaces had lower stiffness in the stress relaxation pullout model than in the standard pullout model, but it was significant in only the cylindrical design group (P<0.05). However, the stress relaxation and standard pullout models did not yield any difference in peak loads in either screw type. Although stress relaxation at the bone-screw interface can alter the mechanical performance of the screw, this may be eliminated by modifying the screw design. A better understanding of viscoelastic properties of the bone-screw interface may help improve implant design and thus, clinical outcomes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.