Abstract

This paper reduces the delay in wireless transmission of electrocardiogram (ECG) data while protecting their essential elements to produce a clinically usable trace. The ECG data are divided into nine layers according to the signal amplitude resolution (6, 8, and 11 bits per sample) and sampling frequency (75, 300, and 2400 Hz). To avoid retransmission, each layer is delivered with a dynamic link adaptation scheme that provides the different levels of error protection in a forward error correction process according to layer importance. This scheme is built on the conservative modulation and coding, as well as on the machine-type communication (MTC) protocol, from LTE-Advanced. Simulations show that our proposed system delivers ECG data that sufficiently support a clinical diagnosis with significantly low latencies (less than 10 ms) under severely erroneous channel conditions.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.