Abstract

This report describes a sampling rate reduction method for heart rate variability monitoring with a wearable device. This work was conducted to realize low-power measurement of biological signals necessary for heart rate variability (HRV) analysis. Continuous operation of the wearable device is an important factor for daily life monitoring. Therefore, the active time of the measuring circuit must be minimized. To reduce the required sampling rate, we propose a sampling error reduction method using interpolation and correlation of the heartbeat waveform. The proposed method is evaluated using measured electrocardiograms from five subjects. Evaluation results demonstrate that the sampling rate can be reduced to 32 Hz with 1 ms RMS error in heartbeat interval and 1.04% LF/HF degradation in HRV analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.