Abstract

Radiofrequency ablation (RFA) is a relatively new technique that has been investigated for the treatment of lung tumors. We evaluated for the first time the in vivo use of saline infusion during radiofrequency ablation of sheep lung. We performed RFA on 5 sheep using open and closed chest RFA and the RITA starburst XL and Xli probes using saline infusion with the Xli probe. The impedance and volume of ablation were compared. A total of 16 ablations were produced, 5 percutaneously and 11 open. The impedance during percutaneous and open RFA without saline infusion was 110 +/- 16.2 and 183.3 +/- 105.8 ohms, respectively. With the saline infusion the impedance was 71.3 +/- 22 ohms and 103.6 +/- 37.5 ohms. The effect of this was significantly larger volume of ablation using the saline infusion during percutaneous RFA (90.6 +/- 23 cm3 vs 10.47 +/- 2.9 cm3, p = 0.01) and open RFA (107.8 +/- 25.8 cm3 vs 24.9 +/- 19.3 cm3, p = 0.0002). Saline infusion during RFA is associated with lower impedance, higher power delivery and larger lesion size.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.