Abstract

To effectively study the mechanisms by which deep brain stimulation (DBS) produces its therapeutic benefit and to evaluate new therapeutic indications, it is vital to administer DBS over an extended period of time in awake, freely behaving animals. To date multiple preclinical stimulators have been designed and described. However, these stimulators have failed to incorporate some of the design criteria necessary to provide a system analogous to those used clinically. Here we define these design criteria and propose an improved and complete preclinical DBS system. This system is fully programmable in frequency, pulse-width and current amplitude, has a rechargeable battery and delivers biphasic, charge-balanced output to two independent electrodes. The system has been optimized for either implantation or for use externally via attachment to rodent jackets.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.