Abstract
An equation was derived to compute the response time of any potentiometric electrode consisting of an inert membrane, an internal solution and a potentiometric sensor. This equation was tested with the CO 2 electrode. The influence of the membrane thickness was studied in particular, and the diffusion coefficient of CO 2 through DuPont de Nemours “Teflon” FEP membranes was determined. Permeabilities of CO 2 and water through these membranes were also measured by other methods. The influence of various factors on the response time of the CO 2 electrode has been discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.