Abstract
The optimization of alpha/beta separation in liquid scintillation using pulse shape analysis is convenient for the simultaneous determination of alpha and beta emitters in natural water and other samples. In this work, alpha/beta separation was studied for different scintillant/vial combinations and it was observed that both the optimum pulse shape discrimination level and the total interference value (that is, the summed relative interference between alpha and beta spectra) were dependent on the sample quenching and independent of the scintillant/vial combination. These results provide a simple method for modifying the counting configuration, such as a change in the cocktail, vial or sample characteristics, without the need to perform exhaustive parameter optimizations. Also, it was observed that, for our counting conditions, the combination of Ultima Gold AB scintillation cocktail with Zinsser low diffusion vials presented the lowest total interference, namely 0.94 ± 0.28%, which is insignificant for the counting of environmental samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.