Abstract

Ten in-vivo failed spinal instrumentation systems, i.e. six Harrington distraction rods with sublaminar hooks, one Harrington distraction rod with segmental wiring and three Luque rods with sublaminar wires, were fractographically analysed. In both Harrington and Luque rods corrosion fatigue was the predominant mechanism resulting in the failure. Five Harrington rods fractured at the first ratchet junction; fractures of the Luque rods were initiated by fretting of sublaminar wire with the rod surface in the presence of spinal non-union. Fretting and crevice corrosion were found to play an important role in compromising the segmentai spinal instrumentation. The susceptibility of the 316 L CW austenitic stainless steel to pitting and relative resistance to crevice corrosion were measured by cyclic anodic polarization tests. The oxide inclusions have been found to play a significant role in the pitting of the alloy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.