Abstract
A roentgen ray fluorescence method was developed for the quantitative determination of iodine concentration in tissue. The method is based on the use of two beams, of which one excites the K roentgen rays of iodine, while the other does not. The concentration of iodine was determined from the difference between the counting rates of the fluorescent and the scattered radiation. The measurements were carried out with standard equipment consisting of a generator, a NaI)Tl)-detector and a scaler. An accuracy of 0.1 mg I per g was obtained with the radiation exposure of about 1 R.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.