Abstract
BackgroundWe investigated the clinical performance of a quantitative multi-modal SPECT/CT reconstruction platform for yielding radioactivity concentrations of bone imaging with 99mTc-methylene diphosphonate (MDP) or 99mTc-dicarboxypropane diphosphonate (DPD). The novel reconstruction incorporates CT-derived tissue information while preserving the delineation of tissue boundaries. We assessed image-based reader concordance and confidence, and determined lesion classification and SUV thresholds from ROC analysis.MethodsSeventy-two cancer patients were scanned at three US and two German clinical sites, each contributing two experienced board-certified nuclear medicine physicians as readers. We compared four variants of the reconstructed data resulting from the Flash3D (F3D) and the xSPECT Bone™ (xB) iterative reconstruction methods and presented images to the readers with and without a fused CT, resulting in four combinations. We used an all-or-none approach for inclusion, compiling results only when a reader completed all reads in a subset. After the final read, we conducted a “surrogate truth” reading, presenting all data to each reader. For any remaining discordant lesions, we conducted a consensus read. We next undertook ROC analysis to determine SUV thresholds for differentiating benign and lesional uptake.ResultsOn a five-point rating scale of image quality, xB was deemed better by almost two points in resolution and one point better in overall acceptance compared to F3D. The absolute agreement of the rendered decision between the nine readers was significantly higher with CT information either inside the reconstruction (xB, xBCT) or simply through image fusion (F3DCT): 0.70 (xBCT), 0.67 (F3DCT), 0.64 (xB), and 0.46 (F3D). The confidence level to characterize the lesion was significantly higher (3.03x w/o CT, 1.32x w/CT) for xB than for F3D. There was high correlation between xB and F3D scores for lesion detection and classification, but lesion detection confidence was 41% higher w/o CT, and 21% higher w/CT for xB compared to F3D. Without CT, xB had 6.6% higher sensitivity, 7.1% higher specificity, and 6.9% greater AUC compared to F3D, and similarly with CT-fusion. The overall SUV-criterion (SUVc) of xB (12) exceeded that for xSPECT Quant™ (xQ; 9), an approach not using the tissue delineation of xB. SUV critical numbers depended on lesion volume and location. For non-joint lesions > 6 ml, the AUC for xQ and xB was 94%, with SUVc > 9.28 (xQ) or > 9.68 (xB); for non-joint lesions ≤ 6 ml, AUCs were 81% (xQ) and 88% (xB), and SUVc > 8.2 (xQ) or > 9.1 (xB). For joint lesions, the AUC was 80% (xQ) and 83% (xB), with SUVc > 8.61 (xQ) or > 13.4 (xB).ConclusionThe incorporation of high-resolution CT-based tissue delineation in SPECT reconstruction (xSPECT Bone) provides better resolution and detects smaller lesions (6 ml), and the CT component facilitates lesion characterization. Our approach increases confidence, concordance, and accuracy for readers with a wide range of experience. The xB method retained high reading accuracy, despite the unfamiliar image presentation, having greatest impact for smaller lesions, and better localization of foci relative to bone anatomy. The quantitative assessment yielded an SUV-threshold for sensitively distinguishing benign and malignant lesions. Ongoing efforts shall establish clinically usable protocols and SUV thresholds for decision-making based on quantitative SPECT.
Highlights
Bone imaging with 99mTc 2,3-dicarboxypropane-1,1-diphosphonate (DPD) or methylene diphosphonate (MDP) is one of the main pillars of nuclear medicine for oncology and orthopedic applications (Buell et al 1982; Lantto et al 1987)
Each patient received a nominal activity of 740 MBq 99mTcMDP or -dicarboxypropane diphosphonate (DPD), depending on the site’s clinical practice, and had given informed consent for their participation in the research studies, which were approved by local investigational review boards (IRB)
We first showed that readers are more consistent and more confident in the xSPECT Bone read as compared to the current clinical state-of-the art method, and we showed that the reads were more correct with xSPECT Bone relative to a consensus-based “truth,” wherein all relevant medical information are available for interpretation
Summary
Bone imaging with 99mTc 2,3-dicarboxypropane-1,1-diphosphonate (DPD) or methylene diphosphonate (MDP) is one of the main pillars of nuclear medicine for oncology and orthopedic applications (Buell et al 1982; Lantto et al 1987). The detection and precise localization of small lesions remains challenging due to the comparatively low SPECT resolution, which is limited to about 8–12 mm by the collimated image formation. We intended to overcome these limitations of bone SPECT by first developing a quantitative reconstruction with a standardized calibration and exploiting anatomical information derived from the CT modality as already installed in contemporary general purpose SPECT/CT systems. We investigated the clinical performance of a quantitative multi-modal SPECT/CT reconstruction platform for yielding radioactivity concentrations of bone imaging with 99mTc-methylene diphosphonate (MDP) or 99mTc-dicarboxypropane diphosphonate (DPD). We assessed imagebased reader concordance and confidence, and determined lesion classification and SUV thresholds from ROC analysis
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.