Abstract
Chiral sum frequency generation spectroscopy (SFG) is of great interest for studying biological systems, among others. Whereas the chiral response in circular dichroism is about 0.1% of the achiral response, the chiral SFG response can be the same order of magnitude as the achiral SFG signal. However, chiral SFG is limited by the attainable signal-to-noise of the weak nonlinear signals and therefore extremely sensitive to proper alignment. We present a robust method for chiral SFG and demonstrate the use on solid-air surfaces with achiral and chiral molecules. We simultaneously measure two orthogonal polarizations-either the interference chiral SFG (±45° polarized) or the pure chiral and achiral SFG-using a waveplate and beam displacer. Both optics are placed in the detection arm and can be easily incorporated into any SFG setup. Furthermore, we employ self-referencing to calibrate alignment for each sample individually using a polarizer in the detection arm. These methods greatly increase the reliability and quality of chiral SFG measurements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.