Object Analyze the adverse events of bakri postpartum balloon in order to promote its safety. Methods Collect and analyze the adverse events from the National Medical Device Adverse Event Monitoring Information System and the U.S. FDA MAUDE database. Chinese and English instructions for use are comparative analysed. Results In most cases, adverse events are balloon rupture and fluid leakage, which attribute to insufficient instruction. The description of the sequence that suturing the myometrium and filling the balloon is deficient. It should be moved gently to avoid puncturing or scratching the balloon with sharp instrument. Conclusion The product instruction improvement as well as training enhancement is recommended for safety consideration. Adverse events monitoring is suggested reinforce simultaneously.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.