Abstract

Several clinical magnetic resonance applications at 1.5 T and higher field strengths require a careful selection of the RadioFrequency (RF) coil design to optimize the RF spatial distribution and sensitivity. Specifically, two basic requirements must be fulfilled for obtaining high quality images: in the transmission mode, RF coils must be able to produce a uniform magnetic field in the volume of interest so that the nuclei can be properly excited; in the receiving mode, a high signal to noise ratio is needed, and the coil must be able to collect the signal emitted by the nuclei with better sensitivity throughout the volume of interest. A number of analytical and numerical methods are reported in the literature to simulate the RF field distributions of surface and volume coils. In this work, we compared the performances of two computational methods (Method of Moments and Finite Elements Method), considering the modelling of both surface (single loop, figure of eight coil - FO8, dual loop) and volume (TEM) coils. Low (1.5 T) and high (7 T) field operating regimes have been analysed. Measurements obtained on the workbench and with a 1.5 T scanner are used for validation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.