Abstract

The articular surface replacement (ASR) total hip arthroplasty (THA) showed accelerated failure rates due to adverse-reaction to metal debris (ARMD). Literature correlating preoperative with intraoperative revision findings respectively post-revision outcome results are rare. 30 of 99 available ASR THA were revised due to ARMD. Mean post-revision follow-up term was 2.3years. In part, preoperative data did not correlate with intraoperative revision findings. ARMD was even found in asymptomatic patients with non-elevated ion levels. Postoperative pain and metal ions decreased significantly (P≤0.016). Cobalt decreased faster than chrome. Patients with intraoperative pseudotumors, osteolysis or bilateral THA did not have higher pre- or postoperative ion values (P≥0.053). Females showed higher postoperative chrome levels (P=0.031). One major post-revision complication (femoral nerve palsy) and one re-revision (late onset infection) occurred.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.