A blade implant that was retrieved in 1990 after 231 months of clinical function (since 1971) was analyzed with respect to clinical, histological, and biomechanical characteristics. The implant clinical records demonstrated no abnormalities or pathological lesions over the tenure of treatment. The bone to implant interface showed a mixture of interfacial tissue components and conditions with adequate direct bone contact (46.4 to 82.3 percent) for classification as osseointegrated. The abutment fracture leading to removal was characterized as a cyclic fatigue mechanism and the distribution of tissue components along the interface could not be correlated with specific biomechanical loading directions. This report considers the clinical and biomechanical records as they relate to the detailed histological investigation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.