Abstract

Response calculations are presented for four commercially available “Knollenberg” light-scattering aerosol counters: the classical scattering aerosol spectrometer probe (CSASP), the active scattering aerosol spectrometer probe (ASASP), the forward scattering spectrometer probe (FSSP) and the axially scattering spectrometer probe (ASSP). The results show sensitivity of the response to aerosol refractive index for values of indexes characteristic of atmospheric aerosols, and for a particular index, multivalued response for particles in the 0.5–4 μm radius range. The response calculations have been validated for two of these instruments (the CSASP and ASASP) by measurement of monodisperse spherical particles. The size resolution of these two instruments is significantly less than advertised by the manufacturer and measurement of irregular particles causes additional loss of resolution.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.