Abstract

The objective of this work was to perform the quantitative analysis of 4 amino acids mixtures: phenylalanine, tyrosine, ornithine and glutamic acid. For this purpose a potentiometric sensor array (electronic tongue) and Sequential Injection Analysis (SIA) measurement system were employed. The flow-through sensor array was composed of 6 miniaturized classical ion-selective electrodes based on polymeric membranes (plasticized PVC) containing 4-octyloxyphenylboronic acid as an ionophore and/or an ion exchanger. Partial Least Squares (PLS) analysis of the steady-state sensor array responses, measured in amino acids mixtures prepared by the SIA system (in two buffer solutions), permitted a correct quantitative analysis only of glutamic acid and ornithine. While a further chemometric treatment, involving the extraction of dynamic components of the transient response employing the Wavelet transform, the removal of less-significant coefficients by means of Causal Index pruning and training of an Artificial Neural Network (ANN) with the selected coefficients, allowed the simultaneous determination of the four amino acids.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.