Abstract

Incomplete charge collection in semiconductor X-ray detectors due to carrier trapping is recognized as an important source of signal broadening. In this paper we show the results of calculations of energy resolution for a TlBr detector using an analytic approach developed in our earlier work in which fluctuations in the distribution of photon absorption sites are related to fluctuations in the collected charge. Using measured values of transport parameters for electrons and holes in the detector material we obtained excellent agreement with experiment in the X-ray energy range 6–660 keV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.