Abstract

There has been significant growth in non-invasive screening techniques for evaluating abnormal tissue. Vibroacoustography (VA), an imaging modality based on ultrasound-stimulated acoustic emission and non-linear scattering characteristics of the target, has previously been used to generate relative real-time, pathology-specific image contrast between abnormal tissue and normal surrounding tissue; however, an in-depth tissue assessment has yet to be completed. VA utilizes two non-destructive low MHz ultrasound tones to produce an acoustic beat frequency in the low kHz range. The acoustic radiation force perturbs the target as a function of its mechanical and acoustic properties and the emissive acoustic waves are detected by a nearby hydrophone to form an image based on the viscoelastic characteristics of the target. We have previously reported that our VA imaging system can distinguish suspect tissue from normal tissue in tissue-mimicking phantoms (TMPs) and ex vivo models with high image contrast; however, the goal of this work is to assess the measurement performance and resolution capabilities of this system in pre-clinical models, specifically TMPs in tissue assessment. Lateral and axial resolution, as well as material characterization, studies were performed on isotropic two-layered and multiple-layered TMP targets. The resolution studies resulted in ∼1 mm lateral and ∼12 mm axial, which were confirmed and validated for a confocal transducer geometry. The system showed sufficient measurement performance to detect regions with elastic moduli difference of at least 10 kPa and lateral width of at least 4 mm. This result, coupled with high imaging contrast, supports the utilization of VA for potential applications in in vivo medical imaging and tissue assessment for intraoperative applications.

Highlights

  • Changes in elasticity of soft tissues are closely related to tissue pathology, as differences in the stiffness of tumor tissue are due to morphological changes from tumorigenesis.[1,2,3] estimation of this mechanical parameter for tumor characterization remains relatively difficult using current intraoperative techniques

  • To first assess the resolution of the VA system, we evaluated the normalized transverse amplitude component of the collected signal from the 1 mm diameter stainless steel bead embedded in the 15% gelatin phantom, at ∆f = 38 kHz

  • The beam profiles of the emitted radiation amplitude for a point target, a 1 mm stainless steel bead embedded in 15% gelatin tissue-mimicking phantoms (TMPs), on the lateral and axial planes, were derived and validated by the experimental findings

Read more

Summary

Introduction

Changes in elasticity of soft tissues are closely related to tissue pathology, as differences in the stiffness of tumor tissue are due to morphological changes from tumorigenesis.[1,2,3] estimation of this mechanical parameter for tumor characterization remains relatively difficult using current intraoperative techniques. Method for tumor margin identification within head and neck surgical subsites.[4,5] Palpation requires considerable clinical expertise and is very difficult to utilize when deeper tissues are involved, as its depth of penetration is low.[6,7] Intraoperative detection of tissue lesions calls for more accurate and sensitive diagnostic tools, as tumor margins remain the highest prognostic indicator for postoperative cancer recurrence.[2,8,9] preoperative imaging techniques have greatly advanced over the past decades, available modalities within the operating room are relatively limited. We further characterized the system’s resolution limits, and examined its ability to produce highly-resolved images of tissue with adequate contrast for applications in medical imaging

Objectives
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.