Abstract

The resolution and long‐term drift of the Advanced Photon Source (APS) beam position monitor (BPM) electronics were measured using the charged particle beams in the ESRF storage ring with various beam current and configurations (single bunch, 8 and 16 equally spaced bunches, and 1/3‐fill). The energy of the stored electrons was 6 GeV. The integrated BPM electronics system as used for this work is capable of measuring the beam position on a turn‐by‐turn basis, which can be accumulated for N turns (N=2n, n=1,2,...,11) . Estimation of the BPM resolution apart from the low‐frequency beam motion was made by measuring the standard deviation in the measured beam position with different Ns. The analysis of the results indicates a BPM resolution of 18/√N [μm] for the APS storage ring, which is equivalent to 0.07 μm/√Hz. For the miniature insertion device BPM with 2.8 times higher sensitivity, the resolution will be 0.02 μm/√Hz. The long‐term drift of the BPM electronics independent of the actual beam motion was me...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.