Abstract

The minimum resistance of expiratory Heidbrink-type value has been assessed by determining the pressure decrease across the values at a flow rate of 30 litre min--1. The average resistance of 70 valves currently in clinical use was 318 Pa at that flow rate and about 44% had resistances greater than the limit of 294 Pa previously suggested. The resistances obtained in new valves were either similar or fractionally greater. Valves being introduced as parts of scavenging systems appeared, on the whole, to have lower resistances.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.