Abstract

The calibration method of the electric gastric lavage machine was developed for the two-way unsteady liquid flow measurement modeland the detection device of the medical electric gastric lavage machine was developed , which flow value can be traced to the quality and time. The experimental results showed that the flow measurement range of the detection device was (50∼3000) mL/min, and the pressure measurement range was (-60∼90) kPa, and the metrological performance of the detection device verified using an active piston type flow standard device can meet the traceability requirements of the key parameters of the medical electric gastric lavage machine.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.