Abstract
The purpose of this study was to evaluate the reproducibility of in vivo quantitative proton MR spectroscopy (MRS) by repeated-measure analysis of variance and intraclass correlation coefficient (ICC) including assessment of both inter- and intrasubject variation. The concentration of metabolites was quantified using long-TR and short-TE stimulated-echo acquisition mode sequences with a quadrature head coil. All preparation procedures for MRS measurement except volume of interest determination were done by an automatic preparation scanning system equivalent to a clinical setting. Analyses were conducted by an LCModel running based on our original basis set by two different reference methods. The ICC showed a good correlation (r = 0.61-0.99) in any metabolite and almost the same value between the two different reference methods, but the concentration of myoinositol had a large intersubject variation. Our results suggest that the reproducibility of quantified metabolite concentration is acceptable in the clinical setting regardless of the reference method, although quantified values may have different intersubject variations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.