Abstract
The reproducibility of data obtained from the recently developed anterior eye segment analysis system (EAS-1000) was evaluated. 40 normal eyes and 62 cataractous eyes were examined at Kanazawa Medical University Hospital or Yayoi Hospital. The radius of the corneal curvature, the corneal thickness, anterior chamber depth, whole lens thickness, anterior chamber angle and the scattering light intensity were all observed on a slit section of the lens, and the intrapupillary area of the cataractous shadow was measured from a retroillumination image. The measurements were made twice at an interval of 14.2 days. There were no significant differences between the data of the 1st and 2nd examination in either place.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.