Abstract

In this reply we discuss the use of two and four-probe methods in the resistivitymeasurements of ITO nanowires. We pointed out that the results obtained byusing two or four probe methods are indistinguishable in our case. Additionallywe present the correct values for resistivity and consequently for the density ofelectrons.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.