Abstract

Aims.Analysis of the prevalence of clinical events and of the trends of the physiologically significant parameters in patients with cardiac implantable electronic devices (CIEDs) with the remote monitoring options.Methods.In 9 clinical centers of the Russian Federation and 2 clinical centers of the Republic of Kazakhstan, 126 patients with an ICD or a pacemaker provided with the Home Monitoring technology (BIOTRONIK, Berlin, Germany) have been enrolled into the ReHoming (Registry Home Monitoring) clinical study. Based on the daily data transmission, all alarm alerts and all the Home Monitoring options changes have been registered with dated alert content and undertaken measures.Results.The study patients, followed up at least for one year, demonstrated 42 adverse events (AE), 26 of which were serious AE (SAE) and 3 SAE were defined as device related (SADE). ICD patients (n=90) had statistically significantly higher SAE prevalence with attendant coronary artery disease (CAD) (p=0.0249). Patients with CRT/D compared to patients with dual-chamber or single-chamber ICD had less SAE rate (р=0.046). Downloads of Home Monitoring parameters for retrospective mathematical analysis were available for 60 ICD patients, 47 of which had episodes of ventricular tachycardia (VT), ventricular fibrillation (VF) and/or atrial tachyarrhythmia (AT). Machine learning analysis of the trends of the physiologically meaningful parameters revealed correlation of the changes with arrhythmia episodes, the random forest method and the gradient boosting method giving the results strongly exceeding a random guess.Conclusion.Home Monitoring of CIED patients enables evaluation of clinical advantages of different device types application, also in regard to prevention of adverse events and possible iatrogenic effects of electrotherapy of the heart. The study results demonstrate a possibility to develop a predictor of arrhythmia episodes, based on daily transmission of trends of physiologically meaningful Home Monitoring parameters.

Highlights

  • In 9 clinical centers of the Russian Federation and 2 clinical centers of the Republic of Kazakhstan, 126 patients with an ICD or a pacemaker provided with the Home Monitoring technology (BIOTRONIK, Berlin, Germany) have been enrolled into the ReHoming (Registry Home Monitoring) clinical study

  • The study patients, followed up at least for one year, demonstrated 42 adverse events (AE), 26 of which were serious AE (SAE) and 3 SAE were defined as device related (SADE)

  • Patients with CRT/D compared to patients with dual-chamber or single-chamber ICD had less SAE rate (р=0.046)

Read more

Summary

Methods

In 9 clinical centers of the Russian Federation and 2 clinical centers of the Republic of Kazakhstan, 126 patients with an ICD or a pacemaker provided with the Home Monitoring technology (BIOTRONIK, Berlin, Germany) have been enrolled into the ReHoming (Registry Home Monitoring) clinical study. Based on the daily data transmission, all alarm alerts and all the Home Monitoring options changes have been registered with dated alert content and undertaken measures

Results
Conclusion
МАТЕРИАЛ И МЕТОДЫ ИССЛЕДОВАНИЯ
ВПС ППС КМП
Нет ХСН Не оценивали Средний ФК
Вторичная профилактика
Анализ эффективности исследования
Клиническая нагрузка по наблюдению пациентов
Количество приступов
Хроническая сердечная недостаточность и приобретенный порок сердца
Число дней до первой терапии
ОБСУЖДЕНИЕ ПОЛУЧЕННЫХ РЕЗУЛЬТАТОВ
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.