Abstract

Near-infrared spectroscopy (NIRS) is a promising non-invasive technique for the continuous monitoring of tissue oxygen delivery. NIRS detects light absorbance of haemoglobin chromophores to determine tissue oxygen saturation (StO2). As skin colour is also determined by the presence of chromophores, it is plausible that NIRS signal quality may be affected by dark skin pigmentation. Tissue saturation in the anterior compartment of the lower leg during isometric contraction was measured using NIRS in 17 volunteers with dark skin pigmentation. Measurements were continued until StO2 was zero percent or until the signal disappeared. The NIRS device failed to register tissue saturation values at some point in nine of seventeen volunteers. This occurred more often in individuals with darker skin. In patients with a dark pigmented skin, NIRS StO2 measurements should be interpreted with caution, as melanin clearly interferes with the quality of the reflected NIRS signal.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.