Abstract

In spite of criticism, monosyllabic discrimination tests are widely employed for hearing aid selection. The current study was designed to investigate the following questions: (1) How reliable is aided speech discrimination of monosyllables in a background of white noise? (2) Can differences among hearing aids be reliably demonstrated by measuring intelligibility of monosyllables in a background of white noise? and (3) Do hearing aids interact with hearing loss? Twenty subjects with mild-moderate sensorineural hearing losses participated in two experimental sessions. Four hearing aids were evaluated using the NU-6 monosyllables in a background of white noise (s/n = +20 dB). Since the standard deviation of the test-retest differences was 6%, differences between aids were not considered significant unless they exceeded 12%. An interaction between subjects and hearing aids was found; that is, the best aid for one person was not the best for all. Measuring monosyllabic word intelligibility in a background of white noise does not reliably identify a single best aid. However, one or more inferior aids were consistently eliminated in 80% of the subjects.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.