Abstract
ObjectiveThe purpose of this study was to evaluate the intraexaminer and interexaminer reliability of image assessment of infrared thermography of the plantar surface of people with diabetes mellitus. MethodsFifty-one participants with diabetes were included. The interexaminer reliability (test) consisted of measuring the temperature of the plantar surface by 2 evaluators who separately performed the analysis of thermographic images. The intraexaminer reliability (retest) consisted of the revaluation of the images by a single evaluator 10 days after the first analysis. ResultsThe analysis of intrareliability and interreliability indicated excellent levels of correlation (intraclass correlation coefficient >0.75, 95% confidence interval >0.70) in the plantar surface points analyzed. ConclusionInfrared thermography image evaluation identified intrareliability and interreliability for temperature analysis of the plantar surface of people with diabetes mellitus. However, further studies need to be conducted to assess validity and its application in health care.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.