Abstract

Having in mind the potential impact of the results presented by Veprev and Muromtsev (2012) [13] on our knowledge of the universe, we reinvestigated the liquid scintillation measurement of the count rate variations of 3H. Making use of the sophisticated Quantulus liquid scintillation spectrometer, we found that the measurement of the high-energy tail of 3H spectrum may be significantly influenced by instrumental instability. Thus, the possible explanation for the relatively high count rate variations of Veprev and Muromtsev (2012) [13] can be attributed mainly to the walk of the cut-off in the integrated spectrum, although weak variations of different origin could be masked by such cut-off drifts. In our experiment we have also registered the oscillatory behavior of measured high-energy tail of 3H spectrum, but with very small amplitude (less than 0.5%), which cannot be easily explained only by instrumental instability. When the total 3H spectrum was measured, no significant variations in the count rate were found.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.