Abstract

We report the gated detection of reflection high-energy electron diffraction (RHEED) signal intensity oscillations during molecular-beam epitaxy (MBE) while the wafer is being rotated at high speed (≳100 rpm). In general, a larger number of oscillation periods are observed with this technique than during stationary measurement because of the very high growth rate uniformity across the sample. We found that the average over all azimuths of the RHEED specular spot intensity shows the same growth induced oscillations. This allowed us to replace the gated detection by an averaging detection method: a low-pass filter suppresses fast variations due to rapid changes in azimuth, while passing low-frequency (<1 Hz) growth related oscillations. This simplifies the detection system and at the same time reduces noise related to wafer motion, 60 Hz, etc. These techniques have potential use for in situ monitoring of thickness and composition of device wafers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.