Abstract

This study was carried out to evaluate the sensitivity and rapidity of the BACTEC radiometric techniques for isolation and susceptibility testing of mycobacteria from extrapulmonary specimens. Concentrated specimens of urine, pleural fluid, and blood as well as other extrapulmonary specimens were processed for the recovery of mycobacteria and for drug susceptibility testing, employing conventional and BACTEC radiometric methods. Out of 483 specimens processed, 20 were found to be positive for Mycobacterium tuberculosis on the conventional Lowenstein -Jensen medium, and 19 were found to be positive in the BACTEC 7H12 medium. Average recovery times were 22.5 days for the conventional method and 10.9 days for the BACTEC method. When isolated cultures were tested for susceptibility to streptomycin, isoniazid, rifampin, and ethambutol, results were reported at an average time of 22 and 5.4 days for the conventional and BACTEC methods, respectively, with good correlation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.