Abstract

Stereotactic burr-hole procedures like biopsies of brain tumours based on CT scan or MRI and angiographic data have so far usually been carried out without real-time ultrasound image control. Intra-operative real-time ultrasound imaging was carried out during twelve target-point stereotactic procedures via a single standard burr-hole using a new slender ultrasound transducer with a diameter of 8 mm. The technical parameters of the transducer are: frequency range of 5 < - > 3.5 MHZ, phased array sector scan, 90 degree sector. The transducer has a bayonet-like configuration and can be sterilized. Sufficient quality of the images was achieved in these twelve cases with different pathological entities such as malformation cysts (3 cases), gliomas (7 cases), one metastasis and one intracerebellar haemorrhage. Moreover, co-ordinate values may be calculated from the ultrasound images generated peroperatively, allowing the surgeon to choose additional targets. Colour flow mapping provides the visualization of vascular structures. For the beginner stereotaxy may be easier to learn using ultrasound real-time imaging.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.