Abstract

We interfaced color Doppler Fourier domain optical coherence tomography (CD-FDOCT) with a commercial OCT system to perform in vivo studies of human retinal blood flow in real time. FDOCT does not need reference arm scanning and records one full depth and Doppler profile in parallel. The system operates with an equivalent A-scan rate of 25 kHz and allows real time imaging of the color encoded Doppler information together with the tissue morphology at a rate of 2-4 tomograms (40 x 512 pixel) per second. The recording time of a single tomogram (160 x 512 data points) is only 6,4ms. Despite the high detection speed we achieve a system sensitivity of 86dB using a beam power of 500microW at the cornea. The fundus camera allows simultaneous view for selection of the region of interest. We observe bi-directional blood flow and pulsatility of blood velocity in retinal vessels with a Doppler detection bandwidth of 12.5 kHz and a longitudinal velocity sensitivity in tissue of 200microm/s.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.