Abstract

Using modeling and simulation, we quantify the influence of spatiotemporal dynamics on the accuracy of data obtained from sensors placed in microscaled reaction volumes. The model refers to cellular reaction (i.e. proton extrusion and oxygen consumption) in complex, buffering solutions. Whole cells or viable tissues cultured in such devices are monitored in real time with integrated sensors for pH and dissolved oxygen. A 3D finite element model of diffusion and metabolic reaction was set up. With respect to pH, the effect of buffering species on proton diffusion is analysed in detail. To account for the delayed time response of real sensors, the sensor impulse response time was implemented by linear convolution. A validation of the model has been achieved by an electrochemical approach. The model reveals significant deviations of measured pH and O2, and values of these parameters actually occurring at different sites of the cell culture volume. It is applicable to any setting of (bio-) sensors involving reaction and diffusion of dissolved gases and particularly H(+) ions in buffered solutions.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.